3-2-2. Slightly pull out the right coarse handle.
3-2-3. Insert the stopper pin that is installed in
left coarse handle A8467300 into the other
circular part of spring AA784200, and push
the left coarse handle in the arrow direc-
tion.
3-2-4. Turn the left coarse handle AB467300
clockwise until it stops, move the left
coarse handle a little away from the body,
and pass it over HK screw AA146300.