7.5 Reprocessing the light source and accessories
99
CLV-190 INSTRUCTION MANUAL
Ch.7
Disinfection
WARNING
Follow the usage (temperature, contact time, use life) and amount (concentration)
provided by surface disinfectant cleaner manufacturer. Failure to follow
manufacturer’s instructions may result in insufficient cleaning and disinfection
effect.
CAUTION
Wipe off remaining surface disinfectant cleaner solution according to instructions of
surface disinfectant cleaner manufacturer. Failure to do so may adversely affect the
human body or this instrument.
After wiping with a moistened lint-free cloth, dry the light source and accessories
thoroughly before using it again. If it is used while still wet, there is the risk of an
electric shock.
1 Turn OFF the light source and disconnect the power cord from the hospital grade wall
mains outlet.
2 Prepare a lint-free cloth moistened with the
surface disinfectant cleaner.
Figure 7.1
3 Wipe off all patient debris from the light source and accessories this instrument using
the moistened lint-free cloth.
Lint-free cloth moistened with
the surface disinfectant cleaner