7.5 Reprocessing the light source and accessories
CLV-190 INSTRUCTION MANUAL
Ch.7
Reprocessing workflow for the light source and accessories
This chapter describes the workflow for reprocessing the light source and accessories.
7.5 Reprocessing the light source and accessories
1 Confirm the appearance of the light source and accessories.
2 If patient debris directly or indirectly adheres to this instrument and accessories, refer
to “ Disinfection” on page 99. Otherwise, be sure to follow the next step.
3 If disinfection is necessary according to national, local, or facility standards, refer to
Disinfection” on page 99. Otherwise, refer to “ Cleaning” on page 100.
3. Disinfection is necessary according to
national, local, or facility standards
2. Patient debris directly or indirectly adheres to this instrument
Cleaning
(Cleaning”)
Disinfection
(Disinfection”)
Yes
Yes
No
No
1. Confirm the appearance of the light source