10
Important Information — Please Read Before Use
CLV-190 INSTRUCTION MANUAL
NOTE
As defined by the international safety standard (IEC 60601-1), medical electrical
equipment is classified into the following types: TYPE CF applied part (the
instrument can safely be applied to any part of the body, including the heart), and
TYPE B/BF applied part (the instrument can safely be applied to any organ except
the heart). The part of the body that an endoscope or electrosurgical accessory can
safely be applied to depends on the classification of the equipment to which the
instruments are connected. Before beginning the procedure, check the current
leakage classification type of each instrument to be used for the procedure.
Classification types are clearly specified in the instruments’ instruction manuals.
Symbol Classification
TYPE CF applied part
TYPE BF applied part
TYPE B applied part