EasyManua.ls Logo

Olympus USPM-RUIII Series - Page 107

Olympus USPM-RUIII Series
143 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
15. USPM Setup Procedure
PV6253-F4E016 65
15.2. Check of PC Card setup
When a USB type control box is in use:
Start up the system through the control panel. Select Device Manager from Hardware. Click “Port
(COM and LPT)”, and ensure that "CP2101 USB to UART Bridge Controller (COM*)" is
registered. ("*" denotes the port No. assigned to the control box.)
When a GP-IB type control box is in use:
Start up the system through the control panel. Select Device Manager from Hardware. Click
“Otherdevice”, and ensure that "REX5052 GPIB PC CARD For DOS/V" is registered.
15.3. Setup of USPM-RU Ver. 3.21
Load the Install CD-ROM in the CD-ROM drive. Click to execute "..¥JPN¥install¥setup.exe" on
the CD-ROM. Select the program to be installed, according to the table.
Japanese version for 380nm
JPN¥install¥setup.exe
English version for 380nm
ENG¥install_en¥setup.exe
Japanese version for 440nm
JPN¥install440¥setup.exe
English version for 440nm
ENG¥install440_en¥setup.exe
The [USPM Setup] window will open. Press the [OK] button to proceed.
On the screen that opens next, select the directory in which the program is to be installed, and
press the button at the upper left. This will initiate the setup. Install the program as instructed by
the messages.
Note: If an error is issued during the installation, press the [Yes] button and continue with the
installation.
When the installation is finished, [USPM] (or USPM440) will be added to the Windows XP
[Start] menu.
When the software is launched for the first time after the installation, an error will be displayed,
since the uspm.int/uspm.env/uspm.cal files do not exist yet. Clicking [OK] will activate the
program.