3.4.4 Output power
The output power specifications were tested under the following conditions:
Compliance voltage set to 100 %
Symmetrical conditions (0°, 120°, 240°) for 3-phase modes
Frequency 50/60 Hz
Resistive load
No amplifier overload
Typical Guaranteed
Compliance voltage (RMS) > 12 V -
Supply voltage range (DC) 6 ... 17 V -
Output power (AC)
3 x 12.5 A
1 x 12.5 A (L-L)
1 x 37.5 A (parallel)
3 x 90 W @ 9 A
1 x 180 W @ 9 A
1 x 250 W @ 24 A
3 x 85 W @ 9 A
1 x 170 W @ 9 A
1 x 200 W @ 24 A
The compliance voltage is the maximum voltage a current source is able to apply in order to drive the
programmed current. The compliance voltage can be configured by the user. Above 8 A, the
compliance voltage is automatically decreased linearly down to 8 V (RMS) at 12.5 A, which limits the
maximum output power.
0
20
40
60
80
100
0 2.5 5 7.5 10 12.5
Output power (W)
Output current (A)
0
20
40
60
80
100
120
140
160
180
200
220
240
260
0 5 10 15 20 25 30 35 40
Output power (W)
Output current (A)
Typ. output power in 3 x 12.5 A configuration Typ. output power in 1 x 37.5 A configuration
CMC 430 User Manual
22
OMICRON

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Omicron CMC 430 and is the answer not in the manual?

Omicron CMC 430 Specifications

General IconGeneral
BrandOmicron
ModelCMC 430
CategoryTest Equipment
LanguageEnglish

Related product manuals