3.15 Environmental conditions
Characteristic Specification
Operating temperature –25 °C ... +50 °C
Storage and transportation –40 °C ... +70 °C
Maximum altitude
Operating
Storage
4000 m
15000 m
Humidity 5 % ... 95 % relative humidity; no condensation
NOTICE
Equipment damage possible.
Do not use the CMC 430 under conditions that may cause or caused condensation inside or on
the outside of the CMC 430. Condensation may occur, for example, if the CMC 430 is moved
from a cold car to a warm and/or humid environment.
3.16
Mechanical data
Characteristic Specification
Weight (mass) < 8.7 kg / 19.2 lbs
Dimensions (w x h x d) 270 x 150 x 380 mm / 10.6 x 6 x 15''
Protection class IP20 according to EN 60529
3.17 Standards
EMC
Emission
Europe
USA
Canada
EN 61326-1, EN 55032 (Class A), EN 61000-3-2/3
FCC Subpart B of Part 15 Class A
CAN ICES-3 (A)/NMB-3(A)
Immunity IEC/EN 61326-1, IEC/EN 61000-4-2/3/4/5/6/8/11
Mechanical tests
Classification
Random vibration
Shock (half - sine)
Free fall
IEC/EN 60721-3-7 (7M2)
IEC/EN 60068-2-64 (1.8 G
rms
/ 30 min / 10 – 500 Hz)
IEC/EN 60068-2-27 (10 g/11 ms, 15 g/11 ms, 30 g/6 ms)
IEC/EN 60068-2-31 (0.1 m)
Certified safety standards
International/Europe
USA
Canada
IEC/EN 61010-1, IEC/EN 61010-2-030
UL 61010-1, UL 61010-2-030
CAN/CSA-C22.2 No. 61010-1, CAN/CSA-C22.2 No. 61010-2-030
CMC 430 User Manual
42
OMICRON

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Omicron CMC 430 and is the answer not in the manual?

Omicron CMC 430 Specifications

General IconGeneral
BrandOmicron
ModelCMC 430
CategoryTest Equipment
LanguageEnglish

Related product manuals