OMNI 4000/7000 Operations and Maintenance Guide Rev F
7-29
Maintenance
20.
Press the Apply HI function key to
complete the span adjustment or the
Abort function key to discard.
21.
After pressing the Apply HI function key,
the Finalize Calibration? screen displays.
a. If you have trouble calibrating the
input channel, reset the channel by
pressing the Defaults function key;
then attempt to calibrate the
channel again.
b. Press the Yes function key to finalize
the calibration of the input channel.
c. Press the No function key to discard
the calibration changes.
22.
Continue calibrating additional input
channels as needed by repeating
these steps.
Output Calibration
To calibrate the 420 mA output channels, follow these instructions:
1.
a. From the main menu of the front
panel, select Utilities and press OK.
b. Select Calibration, press OK.
c. Select Output Calibration.
2.
Press OK to enter the edit mode of the
Selected Output? field.