EasyManua.ls Logo

Optek 156 - Adjusting the Digital Read-Out

Optek 156
49 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Commissioning
- 36 -
optek-Manual--1004-1022-02--156-US-2017-12-11
www.optek.com
9.7 Adjusting the digital read-out
Tool Screw driver
1. Loosen the measuring range plate on the front at both screws.
2. According to the following table, switch the DIP switches 5 and 6 (13) to ON
position (left) or OFF position (right).
Note!
DIP switches 1 to 4 are used for zero point factory setting.
Proceed as follows to set the end value of the digital read-out:
1. Fill the armature with a slightly hazy liquid or dip the probe into a slightly hazy
liquid. You may as well cover the light beam - if the sensor assembly is not
yet installed - so that a stable measuring result is obtained within the set
measuring range.
2. Connect a multimeter to terminals 8 and 9 on the back of the converter. It
displays the corresponding measuring result in mA.
mA values mean the following: 4 mA = 0 % and 20 mA = 100 %
3. Use a screw driver to set the desired end value of the digital read-out at the
potentiometer (14) turning. Proceed according to this example:
mA display = 14.4 mA correspond to a display of 65 %
Adjust the digital read-out to 065. This corresponds to a display from
0 to 100.
Adjust the digital read-out to 325. This corresponds to a display from
0 to 500.
Fig. 24 Potentiometer for digital read-out adjustment
Tab. 6 DIP switch positions
DIP switch Setting of the decimal place in the digital read-out
No decimal place
XXX
1 decimal place
XX.X
2 decimal places
X.XX
5 OFF ON OFF
6OFF OFF ON
1
13
14

Table of Contents

Related product manuals