Assistant Service Manual
OPTIKON 2000
Code 121008S 2-4 Rev. D
Power: ............................... panel or linear control of the U/S power via
system footswitch depression.
U/S Mode: ........................... Linear or Panel; Continuous, Short Pulse, Single
Burst, Multi-Burst, Continuous Burst,
Programmable Emission Mode
U/S Timer: .......................... from 0.00 minutes to 9.59 minutes – Equivalent
Phaco Time display
EQUIPMENT CLASSIFICATION ACCORDING TO
IEC 60601-1
Type of protection against
electric shock: ...................... class I
Degree of protection against electric shock:
Diathermy: .......................... Type BF, floating both at high and low frequencies
U/S: .................................. Type B
Degree of protection against
harmful ingress of water (unit) : . IPX1
Degree of protection against
harmful ingress of water
(footswitch): ........................ IPX8
Degree of safety of application
in the presence of a flammable
anaesthetic mixture: .............. not suitable
DIMENSIONS
Height: .................................. 165 cm
Width: ................................... 60 cm
Depth: ................................... 55 cm
Weight: .................................. 38 Kg
NOTE:
1) Weight and dimensions shown are approximate.
2) Specifications are subject to change without notice.