EasyManua.ls Logo

optris CTlaser LT - Measurement Specifications [LT Models]

optris CTlaser LT
108 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
-20 -
2.4 Measurement specifications [LT models]
LT
LTF
Temperature range (scalable)
-50...975 °C
Spectral range
8...14 µm
Optical resolution
75:1
50:1
System accuracy
1) 2) 3)
±1 °C or ±1 %
±1,5 °C or ±1,5 %
Repeatability
1) 3)
±0.5 °C or ±0.5 %
±1 °C or ±1 %
Temperature coefficient
4)
±0,05 K/ K or ±0,05 %/ K
NETD
3) 5)
0.1 K
0.5 K
Response time (90 % signal)
120 ms
9 ms
Warm-up time
10 min
Emissivity/ Gain, Transmissivity
0.100...1.100 (adjustable via programming keys or software)
Signal processing
Average, peak hold, valley hold (adjustable via programming keys or software)
1)
at ambient temp. 23 ±5 °C; at object temperatures >0 °C (LT) / ≥ 20 °C (LTF); response time=1 s
2)
Accuracy for thermocouple output: ±2.5 °C or ±1 %, whichever is greater
3)
ε =1
4)
for ambient temperatures (head) <18 °C and >28 °C; whichever is greater
5)
at time constant of 200 ms (LT) / 100 ms (LTF) and an object temperature of 25 °C
w ww. . com
information@itm.com1.800.561.8187

Table of Contents

Related product manuals