Technical Data 27-
2.11 Measurement specifications [G5/ G7 models]
G5H
G5H1F
G7
Temperature range (scalable)
250...1650 °C
400...1650 °C
100…1200 °C
Optical resolution
5,0 µm
7,9 µm
System accuracy
1) 2), 3)
70:1
45:1
Repeatability
1) 3)
±1,5 °C or ±1 %
Temperature coefficient
4)
±0,05 K/ K or ±0,05 %/ K
NETD
3) 5)
±0,5 °C or ±0,5 %
Response time (90 % signal)
0,1 K
0,5 K
Emissivity/ Gain
80 ms
10 ms
150 ms
Transmissivity
0,100...1,100 (adjustable via programming keys or software)
Signal processing
0,100...1,100 (adjustable via programming keys or software)
Optical resolution
Mittelwert, MAX, MIN (adjustable via programming keys or software)
1)
at ambient temp. 23 ±5 °C; whichever is greater; response time=1 s
2)
Accuracy for thermocouple output: ±2.5 °C or ±1 %
3)
= 1
4)
for ambient temperatures (head) <18 °C and >28 °C; whichever is greater
5)
at time constant 200 ms and object temperature=start of measurement range + 50 °C