Electrical Installation 73-
4.7.4 Functional inputs (not for CTlaser 4M)
The three functional inputs F1 F3 can be programmed with the CompactConnect software, only.
F1 (digital):
trigger (a 0 V level on F1 resets the hold functions)
F2 (analog):
external emissivity adjustment [0–10 V: 0 V ► = 0.1; 9 V ► = 1; 10 V ► = 1,1]
F3 (analog):
external compensation of ambient temperature/ the range is scalable via software [0–10 V ► -40900 °C/ pre-set range: -20200 °C]
F1-F3 (digital):
emissivity (digital choice via table)
A non-connected input represents:
F1 = High | F2, F3 = Low
[High level: ≥ +3 V…+36 V | Low level: ≤ +0,4 V…–36 V]