After the hold time the signal will drop down to the second highest value or will descend
by 1/8 of the difference between the previous peak and the minimum value during the
hold time. This value will be held again for the specified time. After this the signal will
drop down with slow time constant and will follow the current object temperature.
V----
Setup of Valley hold. If the value is set to 0.0 the display will show --- (function
deactivated). In this mode the sensor waits for ascending signals. The definition of the
algorithm is according to the peak hold algorithm (inverted).