Setup of the Upper limit of the temperature range. The minimum difference between
upper and lower limit is 20 K. The upper limit can only be set to a value = lower limit +
20 K.
Setup of the Lower limit of the signal output. This setting allows an assignment of a
certain signal output level to the lower limit of the temperature range. The adjustment
range corresponds to the selected output mode (e.g. 0-5 V).
Setup of the Upper limit of the signal output. This setting allows an assignment of a
certain signal output level to the upper limit of the temperature range. The adjustment
range corresponds to the selected output mode (e.g. 0-5 V).
Setup of the Temperature unit [°C or °F].
Setup of the Lower alarm limit. This value corresponds to Alarm 1 [►4.7.6 Alarms]
and is also used as threshold value for relay 1 (if the optional relay board is used).
Setup of the Upper alarm limit. This value corresponds to Alarm 2 [►4.7.6 Alarms]
and is also used as threshold value for relay 2 (if the optional relay board is used).