EasyManua.ls Logo

Orbisphere 3600 series - 1.5.10 Troubleshooting

Orbisphere 3600 series
72 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3600 Analyzer for O
2
Measurement—Operator’s Manual 33
3600O2.OM.E9903
1.5.10 Troubleshooting
Orbisphere analyzers are fully factory tested before shipment. Provided they have
been handled properly and are correctly installed, they should be trouble-free.
Here is a list of possible troubles caused by poor handling or identifiable
installation conditions. If you cannot find a solution here, contact your authorized
Orbisphere service technician.
Problem Probable Cause Possible Solution
Sensor won’t
calibrate, even after
thorough servicing
Repeated calibrations
go beyond “expected
limits” of instrument
Re-select membrane from
“Options/Membrane” menu, section 1.4.11.
Then, recalibrate the sensor.
Instrument’s internal
barometric pressure
sensor needs calibration
Recalibrate using “Calibrate/Barom.
Pressure” menu, section 1.3.3.1. Check
against reliable barometer. Do not correct for
sea level!
H
2
S insensitivity option
unnecessarily enabled
Unless you need “H
2
S insensitive” (which
uses different electrolyte) disable from
“Options/Gas” menu, section 1.4.10.
No analog output
signal (portable
instrument)
Portable instrument
requires mains power
for analog output
Connect to appropriate mains power.
“0000” O
2
levels
displayed
“XXXX” selected for
unit lowest range
Choose “X.XXX, XX.XX or XXX.X” from
“Options/Display Units” menu, section 1.4.1.
Shorter-than-expected
sensor operation (in
relatively high DO
2
)
High O
2
concentrations
require more work from
electrochemical sensor
Shut off analyzer when not needed.
Unexpected/incorrect
DO
2
readings
Sample air leak Set flow rate to 100 ml/min; wait until stable.
Slowly increase to 200 ml/min.
If a significant DO
2
decrease is displayed, a
sample air leak is likely.
High residual current Place sensor in dearated sample; wait for low
reading; check concentration against low
measurement limit (see tables in section 3).
If concentration is significantly higher than
low limit, try a sensor service, section 1.5.1.
Instrument and
“moca3600” PC
program are
operational, but do
not work together
(i.e. no data transfer)
Improper PC-to-
instrument connection
Confirm that:
“Options/Serial Output” is Enabled, section
1.4.5;
All 9 pins are available, if using a 25-to-9
pin converter at the PC;
There are no PC COM port/IRQ conflicts.
Alarm relays won’t
operate properly
Normally open /
normally closed state
reversed
Both NO and NC outputs available; confirm
wiring to LEMO (portable instrument) or
terminal block (process instrument)

Table of Contents