3600 Analyzer for O
2
Measurement—Operator’s Manual 49
3600O2.OM.E9903
For portable instruments, a power supply cable, type H05VV-F 3 X 0.75 mm
2
,
which is supplied with the instrument, must be used. The length of the cable
should be three meters maximum and it should have a separable plug (with three
connectors, L+N+PE), without a locking device, to mate with a socket outlet in the
building. The cable and the plug must conform to an appropriate European or
national standard.
10–30 VDC Power Input
10–30 VDC powered portable instruments have a male Binder connector on their
rear panel. It accepts DC power as follows:
Pin 1: DC Return
Pin 2: Earth (Chassis)
Pin 3: Not connected
Pin 4: 10–30 Volts DC
Male Binder receptacle, rear panel of instrument
The user must make the necessary power wiring connections, using the female
Binder plug supplied with your system. User-supplied power cable can have an
outer diameter between 5 and 8 mm.
To disassemble the plug, squeeze the two clips located in the middle of the plug
(position “C” below). This reveals numbered wiring connections inside, to be
matched one-to-one with the diagram above.
Female Binder plug, exploded view
2.1.3.3 Electrical Signal Connections
Connections from a portable instrument to the oxygen sensor, external pressure
sensor, analog recorder output, alarm output, and RS-232 output cables use the
pre-wired LEMO receptacles on the rear panel of the instrument.
3600 series indicating instrument rear panel, portable model
LEMO connector pin assignments are in section 2.1.4, Instrument Connectors.
Instructions for wiring the supplied LEMO plugs are in section 2.1.6, LEMO
Connector Assembly Instructions. Requirements for these user-supplied cables are
in section 2.1.5, User-supplied Cabling Requirements.
Note that you must ground each cable shield properly to assure proper operation.