2.1.4.4 Recorder Output Wiring Identification
Analog output Portable
LEMO-8 pin
Panel/rack mount
terminal strip
position
Wall mount
terminal block
position
User recorder
signal
Gas out Pin 1 J13.3 J6.1 à Gas in
Temperature out Pin 2 J13.2 J6.3 à Temp. in
Pressure out Pin 3 J13.1 J6.5 à Press. in
Range out * Pin 4 J13.4 J6.7 à Range in
Range ground * Pin 5 J14.4 J6.8 ß Range gnd
Pressure ground Pin 6 J14.1 J6.6 ß Press. gnd
Temp. ground Pin 7 J14.2 J6.4 ß Temp. gnd
Gas ground Pin 8 J14.3 J6.2 ß Gas gnd
* Range output is replaced by a second Gas output in certain models (denoted
by “2 gas” in the Instrument Configurations list, section 3)
LEMO-8 diagram—Recorder output
connector, portable instrument
2.1.4.5 Alarm Output Wiring Identification
Alarm relay * Portable
LEMO-6
pin
Panel/rack mount
terminal strip
position
Wall mount
terminal block
position
User alarm
signal
3600 3600M
Relay “C” (High),
normally open (NO)
Pin 1 J14.8 J13.5 J7.6 NO contact
Relay “C” (High),
normally closed (NC)
Pin 2 J14.5 J13.6 J7.5 NC contact
Relay “C” (High),
common
Pin 3 J13.5 J13.7 J7.4 Common
Relay “C1” (Low),
normally open (NO)
Pin 4 J14.7 J14.5 J7.3 NO contact
Relay “C1” (Low),
normally closed (NC)
Pin 5 J14.6 J14.6 J7.2 NC contact
Relay “C1” (Low),
common
Pin 6 J13.6 J14.7 J7.1 Common
* If alarms are configured for “High-High”, the “C1” relay is set for a “High” alarm and both
relays are set for a “High-High” alarm; if alarms are configured for “General”, both relays are set
for an alarm; see section 1.4.3.
LEMO 6 diagram—Alarm relay
connector, portable instrument