42 3600 Analyzer for O
2
Measurement—Operator’s Manual
3600O2.OM.E9903
2. Installation Guidelines
This section should provide you with all the necessary information to prepare for
and install your instrument. Note that additional technical information is included
in section 3. If you still have any questions or encounter any difficulties, contact
your Orbisphere representative.
2.1 Indicating Instrument Installation
The series 3600/3600M indicating instrument is available in several versions, which
allow different means of mounting, power input, and electrical signal connections.
2.1.1 Panel or 19-inch Rack Mount Instrument Installation
2.1.1.1 Instrument Mounting
The standard panel mount instrument configuration is illustrated below.
3600/3600M indicating instrument, panel mount version
Note: Include at least 100mm additional rear clearance for cable protrusion.
You can use the model 32200 panel mount template as a guide for the necessary
panel cutting and drilling. To use, place the template where you wish to mount the
instrument, and mark the drill holes and cut lines. Your mounting hardware
includes a drill bit; use it to drill the 5.4-mm holes required; then cut the 214.5-mm
wide by 125.5-mm high panel hole.
Four self-clinching bolts secure the instrument to the panel. It is easiest to attach
the bottom two bolts first (tightening with the Allen key supplied), guide the
instrument into place, and then fix the top two bolts to finish the procedure. Once
tightened, the self-clinching bolts assure that one person can accomplish future
removal of the instrument.
A retrofit panel mount (model 32201) is also available. This enables users of
existing Orbisphere analyzers to replace their instruments with the 3600/3600M
series, without additional drilling or panel cutting, as shown below.