50 3600 Analyzer for O
2
Measurement—Operator’s Manual
3600O2.OM.E9903
2.1.4 Instrument Connectors
2.1.4.1 Oxygen Sensor Wiring Identification
Sensor signal LEMO-10 pin LEMO-10 pre-wired
terminal strip connection
3600 3600M
Guard Pin 1 J9.9 J11.1
TH1.B Pin 3 J9.7 J11.3
Anode Pin 4 J9.6 J12.3
TH1.A Pin 6 J9.5 J11.2
Cathode Pin 9 J9.8 J12.1
LEMO-10 diagram—Oxygen
sensor connector, all instruments
2.1.4.2 External Pressure Sensor Input Wiring Identification
Sensor signal LEMO-4 pin LEMO-4 pre-wired
terminal strip connection
3600 3600M
Voltage (+) Pin 1 J9.1 J12.4
Ground Pin 2 J9.2 J12.5
Input (+) Pin 3 J9.3 J11.4
Input (-) Pin 4 J9.4 J11.5
LEMO 4 diagram—External pressure
sensor connector, all instruments
2.1.4.3 RS-232 Wiring Identification
device signal
3600 3600M
TXD Pin 1 J13.8 J13.11 J5.1 à RXD
RXD Pin 2 J13.7 J13.10 J5.2 ß TXD
GND Pin 6 J14.9 J14.11 J5.3 ß
à
GND
RS-232 communication parameters: Baud rate: 9600, Stop Bit: 1, Start Bit: 0, Parity: None
LEMO-6 diagram—RS-232 serial
output connector, portable
instrument