EasyManuals Logo

ORTEC Detective-EX-100T User Manual

Default Icon
114 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #104 background imageLoading...
Page #104 background image
ORTEC
®
Detective-EX/DX
®
-100T Portable HPGe Spectrometer — User’s Manual 931040P / 1014
94
! ROI Peak Peak count stops data acquisition when any ROI channel reaches this value
(maximum value of 2
32
!1 counts [4 ×10
9
]).
! Data Overflow Terminates acquisition when data in any channel exceeds 2
31
!1
(2 ×10
9
) counts.
9.1. Electrical and Mechanical
Dimensions Maximum overall dimensions including handle, Ge detector endcap, neutron
detector and shock absorbers.
Detective-EX-100T: 39.4 cm L × 18.3 cm W × 34.9 cm H (15.5 in. × 7.2 in. × 13.1 in.)
Detective-DX-100T: 39.4 cm L × 16.3 cm W × 34.9 cm H (15.5 in. × 6.55 in. × 13.75 in.)
!Weight
Detective-EX-100T: 12 kg (26.3 lb).
Detective-DX-100T: 11.3 kg (24.4 lb).
Maximum Shock Tolerance 20 g. An accelerometer inside the instrument tracks the maximum
shock to which the Detective-EX has been subjected, and a shock in excess of 20 g will void the
warranty.
Ambient Operating Environment 0 to +40°C, at a relative humidity <90% at 35°C, non-
condensing.
Input Power 10 V to 17 V dc 30 W or from auto-sensing ac-powered battery charger.
External Power DC input and battery charge input, MS3112E12-10-S or Bendix PT02E-
12-10S connector.
Internal Battery Rechargeable, nominal 14.4-V lithium-ion battery pack. Contact ORTEC for
the appropriate replacement kit. Charging circuitry and battery management circuitry internal to
the docking station or PAC-II. Battery lifetime 3–5 years. Access door in right side panel of
case. The compartment is semi-sealed and protects against penetration of dirt and moisture. See
Chapter 10 for instructions on replacing the battery.
Battery Life >3 hours at 25°C when HPGe detector is cold. Battery life can be extended indefi-
nitely by the use of optional, external battery belt. The unit is expected to be kept running once
cold.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ORTEC Detective-EX-100T and is the answer not in the manual?

ORTEC Detective-EX-100T Specifications

General IconGeneral
BrandORTEC
ModelDetective-EX-100T
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals