EasyManuals Logo

ORTEC Detective-EX-100T User Manual

Default Icon
114 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #68 background imageLoading...
Page #68 background image
ORTEC
®
Detective-EX/DX
®
-100T Portable HPGe Spectrometer — User’s Manual 931040P / 1014
58
Figure 58. Set Up the Auto-
Calibration Feature.
Figure 59. Enter the Peak Energy
for a Calibration Source.
to return to the calibration settings screen. The new entry will be available on the source
list.
! To use the auto-calibration feature, mark the Enable box by tapping it. To disable auto-
calibration, simply unmark the box.
Set the auto-calibration interval by tapping the Schedule lists and selecting the desired
time of day as well as one of the following intervals:
Daily.
Weekly; choose the Day of the week.
Monthly; tap in the Day of the month field and use the soft keyboard to enter the day.
5.2.1.1. Setup Requirements for Successful Auto-Calibration
For Auto Calibration to run successfully:
! The Auto Calibration checkbox must be marked (enabled).
! The Detective must be in Survey Mode (not in another monitoring mode).
! The detector status must be
Ready
.
! The unit must be properly connected to external power.
! The appropriate source must be in position.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ORTEC Detective-EX-100T and is the answer not in the manual?

ORTEC Detective-EX-100T Specifications

General IconGeneral
BrandORTEC
ModelDetective-EX-100T
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals