EasyManuals Logo

ORTEC Detective-EX-100T User Manual

Default Icon
114 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #35 background imageLoading...
Page #35 background image
931040P / 1014 2. THE DETECTIVE-EX/DX-100
25
Figure 19. Status of Manual
Calibration.
Figure 20. Apply New Calibration
Selection.
Figure 21. Run a New
Background.
! The final step is to take a representative back-
ground measurement. Figure 21 shows the final
calibration screen. Remove the calibration source,
tap the Run a Background button, then refer to
Section 2.12 below.
NOTE A new background measurement is
required after calibration. If you tap
Back without accepting the new back-
ground, only the gamma and neutron
count-rate meters will function;
Identify and Monitor Modes will be
disabled until another you run and
accept a new background.
For more information on calibration, see Sections 5.2.1
and 5.1.3.
2.12. Running A New Background After Calibration
The manual calibration routine deletes the current
background.chn
file. Therefore, the final step
of manual calibration is to run a new background.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ORTEC Detective-EX-100T and is the answer not in the manual?

ORTEC Detective-EX-100T Specifications

General IconGeneral
BrandORTEC
ModelDetective-EX-100T
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals