DSPEC 50
®
and DSPEC 502
®
Digital Gamma-Ray Spectrometer User’s Manual 932502G / 0618
Note that the above responses can be combined to indicate a combination of warnings such as:
%000048081<CR> Amplifier was not pole zeroed nor was high voltage enabled. The
START was attempted anyway.
%000037079<CR> Device already started and high voltage not enabled. The
command was ignored.
%000053077<CR> Device already started, high voltage not enabled and amplifier
was not pole zeroed. The command was ignored.
%000038080<CR> Device preset already exceeded and high voltage not enabled. The
command was ignored.
%000054078<CR> Device preset already exceeded, high voltage not enabled and
amplifier was not pole zeroed. The command was ignored.
Power-Up Alert: Power-up just occurred and the selftest results are:
%001000070<CR> All power-up selftest passed.
%003000072<CR> Battery backed-up data lost.
Command Syntax Errors:
%129001082<CR> Invalid command verb.
%129002083<CR> Invalid command noun.
%129003084<CR> Invalid command verb and noun.
%129004085<CR> Invalid command modifier.
%129005086<CR> Invalid command verb and modifier.
%129006087<CR> Invalid command noun and modifier.
%129007088<CR> Invalid command verb, noun, and modifier.
%129132087<CR> Invalid command (verb, noun, and modifier valid but not
together).
Communication Errors:
%130128084<CR> Command checksum incorrect (only when optional checksum
provided).
Execution Errors:
%131128085<CR> Invalid 1st command parameter.
%131129086<CR> Invalid 2nd command parameter.
%131130078<CR> Invalid 3rd command parameter.
%131132080<CR> Invalid number of command parameters.
%131134082<CR> Invalid device or segment selected.
%131135083<CR> Command not allowed while acquisition in progress.
%131136084<CR> Command not allowed in current mode of operation.
%131137085<CR> Hardware failure detected while processing command.
80