932502G / 0618 7. FIRMWARE COMMANDS AND RESPONSES
ENABLE_ZERO_STABILIZATION
Enables the stabilization of the zero peak. See also DISABLE_ZERO_ STABILIZATION,
SHOW_ZERO_STABILIZATION, and INITIALIZE_ZERO_STABILIZATION.
INITIALIZE
Returns the DSPEC 50 (including flash) to the factory default settings. CAUTION: If using
the DIM connector, turn off the detector bias and physically disconnect the DIM cable
from the DSPEC 50 rear panel before issuing this command.
Non-USB Execution Errors:
The INITIALIZE command simulates a power-down/power-up cycle for the MCB after a
simulated loss of battery-backed-up memory.
Response:
%003000072<CR> MCB Power-up occurred/Memory lost/No selftest errors
(Normal Response for INITIALIZE command).
INITIALIZE_GAIN_STABILIZATION
Resets the gain peak stabilization adjustment to unity (no adjustment). This value is reported
as 2048 by the SHOW_GAIN_ADJUSTMENT command. See also
SET_GAIN_ADJUSTMENT, ENABLE_GAIN_ STABILIZATION, and
DISABLE_GAIN_STABILIZATION.
INITIALIZE_ZERO_STABILIZATION
Resets the zero peak stabilization adjustment to unity (no adjustment). This value is reported
as 2048 by the SHOW_ZERO_ADJUSTMENT command. See also
SET_ZERO_ADJUSTMENT, ENABLE_ZERO_STABILIZATION, and
DISABLE_ZERO_STABILIZATION.
LIST_BLRE
Lists the valid settings of the baseline restore function. See also SET_BLRE.
Response:
BLRE 10.0 100.0 10 100 Legal settings are 10.0 to 100.0 in steps of 1.
LIST_CORR_FLAT
Lists the valid settings of the flattop width. See also SET_CORR_FLAT.
Response:
CORR_FLAT !1.0 .992 0 255 Legal settings are from !1 to 0.992 in
(255 !0 + 1 = 256) steps.
LIST_DISPLAY
Not used by the DSPEC 50; included for backward compatibility.