5.Panel Operation
33
(1) Press the front panel Utility function key, press the Display softkey.
(2) Press the ScrSaver softkey to select On or Off.
(3) At On status, you can set the screen saver time. Turn the knob to adjust
the value on the current cursor, use the / direction key to move
cursor left or right, or use the numeric keypad to enter the parameter and
then select Minute as unit. The screen saver time range is 1 to 999
minutes.
Separator
The user can set the separator of the screen display data.
(1) Press the front panel Utility function key, press the Display softkey.
(2) Press the Separator softkey to toggle between Comma, Space, and
Nothing.
Taking the frequency parameter as an example:
Comma
Space
Nothing
CH1/2 Settings
Load
For either of Out1 and Out2 connector at the front panel, the generator has a
50Ω fixed serial output impendence. If the actual load does not match the
specified value, the voltage level displayed would not match the voltage level
of the component under test. This function is used to match the displayed
voltage with the expected one.
The step to set the CH1 or CH2 load value is as follows:
(1) Press the Utility function key, press the CH1/2 Set softkey.
(2) Press the CH1 Load or CH2 Load softkey, press it again to select High Z
or * ohm ("*" represents a value, the default is 50Ω).
(3) To change the load value, after selecting * ohm, turn the knob to adjust
the value on the current cursor, use the / direction key to move
cursor left or right, or use the numeric keypad to enter the parameter and
then select unit. The load range is 1ohm to 10kohm.
Warning: Setup the correct load for right application.
Align Phase : Select Align Phase in the bottom menu to align the initial phase