SECTION 5 REMOTE OPERATION
84
5.5.3.1 IEEE-488.2 STATUS REPORTING (continued)
IEEE-488.2 STATUS REPORTING COMMANDS
Device Identification query. Returns <PPSC,UPC-MM,vN.NN,1338XX> where
MM is the model number (12 or 32),
N.NN is the firmware version number
XX is the firmware assembly indicating installed options
Clears the ESR and the registers as well as the SCPI :EVENt registers and
ERROR/EVENT queue
Read status byte query. Bit 0 is Power Source SHUTDOWN
Service request enable command
Service request enable query
Standard event status enable
Standard event status enable query
Standard event status register query
Operation complete command
5.5.3.2 SCPI STATUS REPORTING
The following lists the SCPI Commands for Status Reporting from the UPC. Refer to the SCPI 1992
standard for more information. Query commands are explicitly shown in this section for clarity (Figure 5.3).
See (5.2) Conventions regarding SCPI.
Query: system error
Returns one of the following types of messages, with details appended to
it if possible. A maximum of 256 characters may be sent as a response
to this query
-0, "No error" No error detected
-100, "Command error” invalid command or query received
-200, "Execution error” can't execute command
with parameters received
-300, "Device-specific error" UPC not properly configured
-400, "Query error" query aborted
NOTE: The SYSTem:ERRor query clears the corresponding error bit in
the ESR
Query: SCPI version
Returns <1992.0>.
Query: Returns serial number string (1-15 characters)
NOTE: The serial number is also displayed on the display at power up