100
1. Regular inspection and daily maintenance
Regular inspection
To ensure safety, inspect the product regularly, at least once a year. The product wears out and deteriorates
differently depending on the operating frequency and environment. Inspect for loose parts and damage and
the operation of the moving parts regularly.
If you notice an irregularity or want a detailed inspection, contact Paramount Bed (see p. 118) or your
distributor.
Part Operation Inspection item
Bed
Head up operation
* See p. 49
Proper operation can be achieved.
There must not be any strange noises.
Knee up operation
* See p. 49
Proper operation can be achieved.
There must not be any strange noises.
Height adjustment operation
* See p. 50
Proper operation can be achieved.
Theremustnotbeanystrangenoisesorsignicantrattling.
Bed operation buttons
* See p. 60 to 67
All bed operation buttons must work.
CPR
* See p. 70
The head section must be lowered with a CPR lever.
Miscellaneous
There must not be any deformation, damage, or detachment of roller bumpers or
resin parts.
There must not be any peeling of labels, and the text must be legible.
There must not be any scratches or deformation at the mattress deck, frame, or
any other parts.
Board
Attachment state
* See p. 29
The board stopper must be locked properly.
Theremustnotbeanysignicantrattling.
Caster
Traveling
There must not be any strange noises when the bed is traveling.
There must not be any abnormal vibrations when the bed is traveling.
Caster locking operation
* See p. 34
Proper operation can be achieved.
Theoperationmustnotbeoverlydicult.
The bed must not move when casters are locked.
The bed must move smoothly when the lock is released.
Side-rail
Side-rail operation
* See p. 30 and 31
Proper operation can be achieved.
Theremustnotbeanystrangenoisesorsignicantrattling.
Theoperationmustnotbeoverlydicult.
Electrical components
Visual check of electrical components
Cables must not get tangled in the bed.
The plug and the cable must have no scratches or damage.
The side-rail integrated panel, patient control panel, and handheld controller (stan-
dard or accessory), BED Navi (accessory) must have no scratches or damage.
The pins of the bed side connector (37-pin male) must not be damaged.
Nursecallnoticationcheck(only
when using the nurse call auxiliary
button of the bed)
The LED lamp above the nurse call auxiliary button must turn on and a nurse call
must be made when the nurse call auxiliary button is pressed while the bed is
connected to the nurse call wall plate.
5 REFERENCE INFORMATION