EasyManua.ls Logo

PARAMOUNT BED 7B01001000A3 - Emergency Operation Procedures

PARAMOUNT BED 7B01001000A3
120 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5
4
1
2
3
69
2 FUNCTIONS AND OPERATIONS
Using the handheld controller (PA-H 612*, PA-H 614* only)
You can adjust the angle of the head and knee sections and the bed height with no increments with the
handheld controller. The handheld controller is intended to be operated by patients.
WARNING
Do not allow it to be operated by a person deemed incapable of
understanding the operation (a child aged 12 or under or a person with
dementia). If a person deemed incapable of understanding the operation
operates the handheld controller by oneself, the body may get caught in
the bed by mistake, resulting in accidental injury.
<Part names and functions>
g Operation buttons
To raise the part to operate, press the button, and
to lower it, press the button.
1 Head and knee coordination buttons
............................................. Adjust the angle of
the head/knee
sections.
2 Head up/down buttons ......... Adjust the angle of
the head section.
3 Knee up/down buttons ......... Adjust the angle of
the knee section.
4 Height up/down buttons ...... Adjust the bed
height.
Note!
n Press the operation button to operate
the bed and release the button to stop
the operation at its position.
n Press two or more operation buttons
simultaneously to stop the operation.
n If you stopped the operation by press-
ing two or more buttons simultaneous-
ly, remove your hands from all buttons
to perform the operation again.
g Indicator lamps
5 Power lamp .......................... Indicates the power
supply status for the
bed.
On (green) : Normal power on status.
Off :
The power plug is disconnected.
Flashing (green) : Indicates the battery status.
See p. 44.

Table of Contents