114
4. Technical explanation regarding EMD (Electromagnetic Disturbance)
The bed requires special safety precautions regarding EMC to be observed. Install, operate, and use the
bed according to the information provided below.
The bed is intended for use in an electromagnetic environment specified according to “Electromagnetic
interference.” Using the bed in an unintended environment may result in the occurrence of the following
problems.
The function for supporting the patient’s body weight may not be maintained.
The emergency head-down function may not be maintained.
No. Cable Maximum length [m] Remarks
1 AC power cable 3.5 Without shield
2
Handheld controller cable of type with a handheld
controller
1.2 Without shield
3
Handheld controller cable
(RC-1850E: accessory)
0.6 Without shield
The bed is intended for use in doctor’s offices, clinics, exclusive medical facilities, independent surgery
centers, multiple treatment facilities, hospitals (except near high-frequency surgical equipment, and outside
RF shielded rooms for magnetic resonance images). The customers or users of the bed must ensure that
the bed is always used in this environment.
Due to its emission characteristics, the bed is suitable for use in hospitals. If used in a home environment,
this device may not offer sufficient protection against radio frequency communication services.
It may be necessary for the user to take relaxation measures such as re-arranging or changing the direction
of the device.
WARNING
n If accessories, transducers, or cables other than those specified and
sold by the manufacturers of the bed are used as replacement parts for
internal components, it may result in an increase in emission or a
decline in immunity of the bed.
n Use portable RF communication devices (including antenna cables and
peripheral devices such as external antennas) at a distance of at least
30 cm from all parts of the bed. Using the RF communication devices
within a distance of 30 cm may cause a decline in the performance of
the bed.
n The bed cannot be used adjacent to another device or in a stacked
state. If the bed must be used adjacent to another device or in a stacked
state, make sure the bed operates appropriately at the location of use.
Guidance and manufacturers declaration — electromagnetic emissions
Thebedisintendedforuseintheelectromagneticenvironmentsspeciedbelow.
Thecustomerortheuserofthebedmustuseitinanenvironmentspeciedhere.
Emission test Compliance Electromagnetic environment — guidance
RF emissions
CISPR 11
Group 1
The bed uses RF energy only for its internal functions.
Therefore, its RF emissions are low and are not likely to
cause any interference with nearby electronic equipment.
RF emissions
CISPR 11
Class A
Use the bed in a power supply network of a professional
environment, such as a hospital or other medical facility.
Harmonic emissions
IEC 61000-3-2
N/A
Voltageuctuations/ickeremissions
IEC 61000-3-3
N/A