Guidance and manufacturers declaration — electromagnetic immunity
The bed is intended for use in the following electromagnetic environment. The customer or user of the bed must use it in
anenvironmentspeciedhere.
Immunity test
IEC 60601-1-2
Test level
Compliance level
Electrostatic discharge (ESD)
IEC 61000-4-2
Contact discharge: ±8 kV
Air discharge: ±2 kV, ±4 kV,
±8 kV, ±15 kV
Contact discharge: ±8 kV
Air discharge: ±2 kV, ±4 kV,
±8 kV, ±15 kV
Electrical fast transient/ burst
IEC 61000-4-4
Power supply line: ±2 kV
Input/output lines: ±1 kV
Power supply line: ±2 kV
Input/output lines: ±1 kV
Surge
IEC 61000-4-5
Line to line: ±0.5 kV, ±1.0 kV
Line to earth: ±0.5 kV, ±1.0 kV, ±2.0 kV
Line to line: ±0.5 kV, ±1.0 kV
Line to earth: ±0.5 kV, ±1.0 kV, ±2.0 kV
Voltage dips, instantaneous power
failure,andvoltageuctuationinpower
supply cable
IEC 61000-4-11
0.5 cycle for test level 0% UT
(Phase: at 0°, 45°, 90°, 135°, 180°,
225°, 270° and 315°)
0.5 cycle for test level 0% UT
(Phase: at 0°, 45°, 90°, 135°, 180°, 225°,
270° and 315°)
1 cycle for test level 0% UT
and
25/30 cycles for test level 70% UT
(Single phase: at 0°)
1 cycle for test level 0% UT
and
25/30 cycles for test level 70% UT
(Single phase: at 0°)
Short-term power failure
IEC 61000-4-11
250/300 cycles for test level 0% UT 250/300 cycles for test level 0% UT
Power frequency
(50/60Hz)magneticeld
IEC 61000-4-8
30 A/m 30 A/m
Note: UT is the AC power voltage before applying the test level.