WARNING
<Using the bed>
g Use the bed with the boards attached, except when providing treatment.
Also, check the patient’s condition when detaching the board.
n A body part may get caught in a gap between the board and other objects, such as a suitable product
or peripheral device, resulting in injury.
n The patient may fall off the bed, resulting in injury.
g Do not use the bed for patients with a height of 146 cm or less, a weight of
40 kg or less, and a BMI of 17 or less.
n This product is not designed for patients with a
height of 146 cm or less, a weight of 40 kg or less,
and a BMI of 17 or less. A part of the body may
get caught in a gap between the bed and side-rail,
resulting in injury.
n Even when the side-rails are used, an infant or
small child may fall off the bed through a gap
between the bed and side-rail, resulting in injury.
g Do not stand on the bed.
The patient may fall off the bed resulting in injury or
the bed may be damaged or deformed. In particular,
be careful with children.
g Do not spill liquids on the bed.
n Do not spill liquids on the bed’s electrical components, such as the connector
*)
, side-rail integrated
panel, and patient control panel. The bed may malfunction resulting in injury or may be damaged due
to a short circuit.
n If liquid is accidentally spilled on the bed, disconnect the power plug from the outlet and wipe off the
liquid. Dry the bed completely and make sure that the bed operates safely before use.
*) Connector to connect optional electrical components, such as a handheld controller, nurse call relay
unit, or Nemuri SCAN
g Do not place the bed in a strong electromagnetic environment such as an
MRI.
n The bed may operate abnormally, resulting in injury or in damage to the bed.