77
Note!
n You cannot mute the alarm sound in
the following cases.
• Whenthe heightfrom oorto mat-
tress deck top has reached 43 cm
and the operation is stopped when
lowering the bed height.
• When you are lowering the bed
height again after pausing the oper-
ation.
• During electric CPR operation.
n The pause height and minimum height
vary depending on the inclination of
the bed.
1. Press the settings button.
2. Press the function selection button for the setting
you want to change.
The setting selection lamp is switched.
3. Press the settings button.
This completes the setting.
g Setting procedure
WARNING
When the setting is completed, press the settings button on the back of
the handheld controller again to confirm the setting. If you do not press
the settings button, the changed setting is not confirmed and you may be
injured by unexpected movement of the bed.
Note!
Settingsspeciedusingfunctionselectionbuttonsandinthenoticationfunctionarenot
canceled even after a power failure or if the power plug or handheld controller plug is discon-
nected.
Press the settings
button.
Press the settings
button.
1
2
3
Speed switching
button
Operation sound
button