EasyManua.ls Logo

Parker 598P - 598;599 P Set-Up Procedure; Calibration Switch; Armature Current Calibration; Field Current Calibration

Parker 598P
26 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
10 of 20
598P/599P Application Manual
598/599P Set-up Procedure
At the moment the software identifies the 590P External Stack Module as a 590PH Module and therefore the
External Stack must be set for a similar current range to an equivalent 590PH.
CALIBRATION SWITCH
1 2
3
4 5 6 7 8 9 10
500 amp Steps
If Cal
V a Cal
ON
4 Sw i t ches On
ARMATURE CURRENT CALIBRATION
4 Switches must be “On” under ALL circumstances. The Drive Software will then scale the Current Feedback
Signal to give the desired Output Current provided the correct ACCTs of a ratio of 2000 to 1 have been fitted.
FIELD CURRENT CALIBRATION
SW7
SW8
SW9
Field Calibration
0
0
0
20
0
0
1
40
0
1
0
60
0
1
1
80
1
0
0
100
1
0
1
120
ARMATURE VOLTAGE CALIBRATION
SW10
Attenuation
Calibration
0
Va/100
Normal
1
Va/200
Va/2
Example
Set the Product Code of the Module to a 590H of the same Current Range as the Stack and with the associated
maximum field current calibration. i.e. for a 2Q Stack with a 2000 Amps Armature and a 35 Amps Field set the
Product Code to DC 2Q 2200A 40 D
Set the Field Calibration switches to the appropriate range i.e. with SW7 OFF, SW8 OFF & SW9 ON.
Adjust the Armature Current and other parameters as normal provided the external CT is 2000:1.
For a different CT Ratio “R” set the Armature Current using an appropriate multiplier where the multiplier M =
2000/R and the Armature Current Setting is M * Normal. i.e. if the CT Ratio is 1000:1 you will need to set 2 *
the required current. Note this may mean that you may have to select a Higher Current Range on the Initial
Product.
This manual was downloaded on www.sdsdrives.com
+44 (0)117 938 1800 - info@sdsdrives.com

Related product manuals