In two samples simultaneously into the two slots reference solution, and then
press the [ZERO] key, the instrument will be blank correction in the current
wavelength. Display calibration is completed 100.0% T or 0.000Abs.
2.5. Measurement data
Calibrated with reference solution 100% T/0Abs, remove the rear slot reference
sample solution into the sample solution, and then press the [START] button,
then perform a test, the sample data are immediately added to the list.
2.6. Delete data
Delete files Press [CLEAR] button, delete files prompt box will pop up, this
operation will delete all the test data currently under test. Select [Yes], the file
will be deleted all the data, select [No] will return the test window.
2.7. Save the file
Press the Save button to save the current list of test data to a file. If this is the
first time you save a file, the dialog box will pop up asking to save the file name
input.