Operation
6.3.2 Rechargeable battery
The battery is protected against deep discharge,
overcharging, overheating and short circuits by
Electronic Cell Protection (ECP). In the event of a
hazard, a protective circuit switches the battery off
automatically. If a fault is detected in the battery,
the LEDs flash on the battery level indicator.
Description Method of resolution
Code:
If the battery is outside its
charging temperature range,
three LEDs will flash on the
battery level indicator.
1 Disconnect charger from
battery.
2 Leave battery to cool
down.
3 Contact your specialist
dealer if the problem
persists.
Code:
If a fault is detected in the
battery, two LEDs flash on the
battery level indicator.
Contact your specialist
dealer.
Code:
No LEDs will flash if the charger
is defective and will not charge.
One or more LEDs will light up
permanently, depending on the
battery level.
Contact your specialist
dealer.
Code:
No LEDs will light up if there is
no power.
1 Check all plug
connections.
2 Check contacts on the
battery for dirt. Clean the
contacts carefully if
needed.
3 Contact your specialist
dealer if the problem
persists.
Table 43: List of battery error messages