EasyManuals Logo

Pendulum CNT-91R Programmer's Handbook

Default Icon
264 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #136 background imageLoading...
Page #136 background image
Command Reference 8-28
:CONFigure :<Measuring Function>
[8 <parameters>[,(<channels>)]]
Configure the counter for a single measurement
Use the configure command instead of the measure query when you want to
change other settings, for instance, the input settings before making the measure
-
ment and fetching the result.
The :CONFigure command controls the settings of the Input, Sense and Trigger sub
-
systems in the counter in order to make the best possible measurement. It also
switches off any calculations with :CALC:STATE 8 OFF.
:READ? or :INITiate;:FETCh? will make the measurement and read the resulting
measured value.
Since you may not know exactly what settings the counter has chosen to configure
itself for the measurement, send an *RST before doing other manual set up mea
-
surements.
Parameters
<Measuring Function>, <Parameters> and <Channels> are defined on page 8-54.
The optional parameter :SCALar means that one measurement is to be done.
Returned format: <String>¿
<String> contains the current measuring function and channel. The response is a
<String data element> containing the same answer as for [:SENSe]:FUNC-
tion?.
Example:
SEND®
:CONF:FREQ:RAT8(@3),(@1)
Configures the counter for freq. ratio C/A.
See also: ‘Explanations of the Measuring Functions’ starting on page 8-59.
Complies with standards: SCPI 1991.0, confirmed.

Table of Contents

Other manuals for Pendulum CNT-91R

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Pendulum CNT-91R and is the answer not in the manual?

Pendulum CNT-91R Specifications

General IconGeneral
BrandPendulum
ModelCNT-91R
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals