8-58 Command Reference
:MEASure :MEMory<N>?
Memory Recall, Measure and Fetch Result
Use this command when you want to measure several parameters fast.
:MEAS:MEM1? recalls the contents of memory 1 and reads out the result,
:MEAS:MEM2? recalls the contents of memory two and reads out the result etc.
The equivalent command sequence is *RCL1;READ?
The allowed range for <N> is 1 to 9. Use the somewhat slower
:MEAS:MEMory?8N command described below if you must use memories 10 to
19.
Returned format:
<measurement result>¿
Complies with standards: SCPI 1991.0, confirmed
:MEASure :MEMory?
8 <N>
Memory Recall, Measure and Fetch Result
Same as above command but somewhat slower. Allows use of all memories (1 to
19).
Example:
:MEAS:MEM 8 13
This example recalls the instrument setting in memory number 13, makes a meas
-
urement, and fetches the result.
Complies with standards: SCPI 1991.0, confirmed