EasyManuals Logo

Pendulum CNT-91R Programmer's Handbook

Default Icon
264 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #204 background imageLoading...
Page #204 background image
8-96 Command Reference
:FREQuency :POWer :UNIT CNT-90XL
8 DBM|W
Input C Measurement Unit
Selects dBm or W as the basic measurement unit to be displayed or read out.
Parameters: DBM | W
The reference level 0 dBm is 1 mW in 50 W. Increasing the level by 3 dB means doubling the
power. Decreasing the level by 3 dB means halving the power.
Returned format:
DBM | W ¿
*RST condition: DBM
:FREQuency :RANGe :LOWer
8 «<Numeric value>|MIN|MAX»
High Speed Voltage Measurements
Use this command to speed up voltage measurements and Autotrigger functions
when you don’t need to measure on low frequencies.
Time to determine trigger levels (typical)
Min. frequency limit
(1 Hz)
Default
(100 Hz)
Max. frequency limit
(50 kHz)
8 s 80 ms 20 ms
Parameters:
<Numeric value> between 1 and 50000 (Hz)
MIN gives 1 Hz
MAX gives 50 kHz
Returned format:
<Numeric value>¿
*RST condition: 100 (Hz)
Complies with standards: SCPI 1991.0, confirmed.

Table of Contents

Other manuals for Pendulum CNT-91R

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Pendulum CNT-91R and is the answer not in the manual?

Pendulum CNT-91R Specifications

General IconGeneral
BrandPendulum
ModelCNT-91R
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals