8-124 Command Reference
:TRIGger :COUNt
8 «<Numeric value> | MIN | MAX»
No. of Triggerings on each Ext Arm start
Sets how many measurements the instrument should make for each ARM:STARt
condition, (block arming).
These measurements are done without any additional arming conditions before the
measurement. This also means that stop arming is disabled for the measurements
inside a block.
+
The actual number of measurements made on each INIT equals to:
(:ARM:START:COUN)*(:TRIG:START:COUNT)
Parameters:
<Numeric value> is a number between 1 and 16777215 (2
24
-1).
MAX gives 16777215
MIN gives 1
Example:
SEND®
:TRIG:COUN 8 50
Returned format: <Numeric value>¿
*RST condition: 1
Complies with standards: SCPI 1991.0, confirmed.
:TRIGger :SOURce
8TIMer | IMMediate
Pacing
Enables or disables the pacing function, i.e. the sample rate control. The pacing
time is set by the :TRIG:TIM command.
Parameters:
TIMer - enables pacing
IMMediate - disables pacing
*RST condition: IMM

Table of Contents

Other manuals for Pendulum CNT-91R

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Pendulum CNT-91R and is the answer not in the manual?

Pendulum CNT-91R Specifications

General IconGeneral
BrandPendulum
ModelCNT-91R
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals