EasyManuals Logo

Pendulum CNT-91R Programmer's Handbook

Default Icon
264 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #189 background imageLoading...
Page #189 background image
Command Reference 8-81
:MEMory :DATA :RECord :SAVE
8 <Dataset Number>[,<Label>]
Save Measurement Data Array to Internal FLASH Memory
One of the eight (0-7) memory positions must be entered, but you can also enter
an optional name (max 6 characters) for easier recognition.
A default name will be assigned automatically if you omit the <Label> parameter. It
represents the abbreviated measurement function and the channel. For example:
Period Single A will read PerA.
If the instrument is not in Hold when this command is sent, then Execution Error
(-200) will be placed in the error queue.
If the instrument is not in Statistics Mode when this command is sent, then Settings
Conflict Error (-221) will be placed in the error queue.
If specified <Dataset> already contains data, then Directory Full Error (-255) will be
placed in the error queue.
If there are more than 32000 samples to save, only the last 32000 will be saved
without notification to the operator.
:MEMory :DATA :RECord :SETTings?
8 <Dataset Number>
Recall Instrument Settings Used for Specified Dataset
The command returns the instrument settings used when the specified <Dataset>
was saved. The format is the same as for :SYSTem:SET.

Table of Contents

Other manuals for Pendulum CNT-91R

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Pendulum CNT-91R and is the answer not in the manual?

Pendulum CNT-91R Specifications

General IconGeneral
BrandPendulum
ModelCNT-91R
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals