8-44 Command Reference
:INITiate
Initiate Measurement
The :INITiate command initiates a measurement. Executing an :INITiate
command changes the counter’s trigger subsystem state from “idle-state” to
“wait-for-bus-arm-state” (see Figure 6-11). The trigger subsystem will continue to
the other states, depending on programming. With the *RST setting, the trigger
subsystem will bypass all its states and make a measurement, then return to idle
state. See also ‘How to use the Trigger Subsystem’ at the end of this chapter.
Complies with standards: SCPI 1991.0, confirmed.
:INITiate :CONTinuous
8 <Boolean>
Continuously Initiated
The trigger system could continuously be initiated with this command. When Con-
tinuous is OFF, the trigger system remains in the “idle-state” until Continuous is set
to ON or the :INITiate is received. When Continuous is set to ON, the comple
-
tion of a measurement cycle immediately starts a new trigger cycle without entering
the “idle-state”, i.e., the counter is continuously measuring and storing response
data.
Returned format: 1|0¿
*RST condition: OFF
Complies with standards: SCPI 1991.0, confirmed.

Table of Contents

Other manuals for Pendulum CNT-91R

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Pendulum CNT-91R and is the answer not in the manual?

Pendulum CNT-91R Specifications

General IconGeneral
BrandPendulum
ModelCNT-91R
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals