8-56 Command Reference
:MEASure :<Measuring Function>?
[8 [<parameters>][ ,(<channels>)]]
Make one measurement
The measure query makes a complete measurement, including configuration and
readout of data. Use measure when you can accept the generic measurement
without fine tuning.
+
When a CONFigure command or MEASure? query is issued, all counter settings
are set to the *RST settings., except those specified as <parameters> and
<channels> in the CONFigure command or MEASure? query.
The :MEASure? query is a compound query identical to:
:ABORt;:CONFigure:<Meas_func>;:READ?
Parameters:
<Measuring Function>, <Parameters> and <Channels> are defined on page 8-54.
You may omit <parameters> and <Channels>, which are then set to default.
Returned format: <data>¿
Where: The format of the returned data is determined by the format commands: :FORMat and
:FORMat:FIXed.
Example:
SEND®
:MEAS:FREQ? 8 (@3)
READ¬ 1.78112526833E+009
This example measures the frequency on the C-input and outputs the result to the
controller.
Type of command: Aborts all previous measurement commands if *WAI is not used.
See also: ‘Explanations of the Measuring Functions’ starting on page 8-59.
Complies with standards: SCPI 1991.0, confirmed.

Table of Contents

Other manuals for Pendulum CNT-91R

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Pendulum CNT-91R and is the answer not in the manual?

Pendulum CNT-91R Specifications

General IconGeneral
BrandPendulum
ModelCNT-91R
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals