Command Reference 8-51
:INPut«[1]|2» :SLOPe
8 «POS|NEG»
Trigger Slope
Selects if the counter should trigger on a positive or a negative transition. Selecting
negative slope is useful for Time Interval measurements.
The slope is fixed for Pos/Neg Pulse Width/Duty Factor and Rise/Fall Time.
Arming slope is not affected by this command. Use :ARM:STARt:SLOPe and
:ARM:STOP:SLOPe instead.
Returned format:
POS | NEG ¿
Example for Input A (1)
SEND®
:INP:SLOP 8 POS
Example for Input B (2)
SEND®
:INP2:SLOP 8 NEG
*RST condition POS
Complies with standards: SCPI 1991.0, confirmed.