EasyManuals Logo

Pendulum CNT-91R Programmer's Handbook

Default Icon
264 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #208 background imageLoading...
Page #208 background image
8-100 Command Reference
:ROSCillator :SOURce
8 «INT|EXT|AUTO»
Select Reference Oscillator
Selects the signal from the external reference input as timebase instead of the in
-
ternal timebase oscillator. If the parameter is set to the default value AUTO,
external reference will be used, if present.
Returned format: <INT|EXT|AUTO>¿
*RST condition: AUTO
Complies with standards: SCPI 1991.0, confirmed.
:TIError :FREQuency :AUTO CNT-91
8 «ON|OFF»
Automatic Recognition of Basic Frequency for TIE Measurement
If AUTO is ON, a check measurement is made at the start of the block to deter-
mine if the frequency of the input signal, rounded to 4 significant digits, is listed for
automatic recognition, for instance:
4, 8, 15.75, 64 kHz or
1.544, 2.048, 5, 10, 27, 34, 45, 52 MHz
If the command is successful, the found value will be stored and can be recalled
with a query command. Subsequent TIE measurements will use this value until it is
changed by sending this command once more or by sending the setting command
:TIError:FREQuency 8 <Numeric value>, which will deliberately fix the fre
-
quency.
Returned format: 1|0¿
*RST condition: OFF¿

Table of Contents

Other manuals for Pendulum CNT-91R

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Pendulum CNT-91R and is the answer not in the manual?

Pendulum CNT-91R Specifications

General IconGeneral
BrandPendulum
ModelCNT-91R
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals