Command Reference 8-137
*PUD
8 Arbitrary block program data>
Protected User Data
Protected user data. This is a data area in which the user may write any data up to
64 characters. The data can always be read, but you can only write data after un
-
protecting the data area. A typical use would be to hold calibration information, us
-
age time, inventory control numbers, etc.
The content at delivery is: #234 FACTORY CALIBRATED ON: 19YY-MM-DD
YY = year, MM = month, DD = day
Returned format: <Arbitrary block response data>¿
Where:
<arbitrary block program data> is the data last programmed with *PUD.
Example
Send ®
:SYST:UNPR; *PUD 8 #240Calibrated 8 1993-07-16, 8 inven
-
tory 8 No.1234
# means that <arbitrary block program data> will follow.
2 means that the two following digits will specify the length of the data block.
40 is the number of characters in this example.
Complies with standards: IEEE 488.2 1987.
*RCL
8<Decimal data>
Recall
Recalls one of the up to 20 previously stored complete instrument settings from the
internal nonvolatile memory of the instrument.
Memory number 0 contains the power-off settings.
Parameters:
<Decimal data> = a number between 0 and 19.
Example:
SEND®
*RCL 8 10¿
Complies with standards: IEEE 488.2 1987.

Table of Contents

Other manuals for Pendulum CNT-91R

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Pendulum CNT-91R and is the answer not in the manual?

Pendulum CNT-91R Specifications

General IconGeneral
BrandPendulum
ModelCNT-91R
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals