EasyManuals Logo

Pendulum CNT-91R Programmer's Handbook

Default Icon
264 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #85 background imageLoading...
Page #85 background image
n
Operation Status Group
This group reports the status of the coun
-
ter measurement cycle.
Bit 8 (weight 256) Measurement
Stopped (MSP)
This bit shows that the counter is not mea-
suring. It is set when the measurement, or
sequence of measurements, stops.
Bit 6 (weight 64) Wait for Bus Arm
-
ing (WFA)
This bit shows that the counter is waiting
for bus arming in the arm state of the trig
-
ger model.
Bit 5 (weight 32) Waiting for Trigger
and/or External Arming (WFT)
This bit shows when the counter is ready
to start a new measurement via the trigger
control option (488.2), for the shortest
possible trigger delay. The counter is now
in the wait for the trigger state of the trig
-
ger model.
Bit 4 (weight 16) Measurement
Started (MST)
This bit shows that the counter is measur
-
ing. It is set when the measurement or se
-
quence of measurements starts.
n
Summary, Operation Status
Reporting
:STAT:OPER:ENAB
Enable reporting of Operation Status in
the status byte.
*SRE 128
Enable SRQ when operation status has
someting to report.
:STAT:OPER?
Reading and clearing the event register of
the Operation Status Register structure
:STAT:OPER:COND?
Reading the condition register of the Op-
eration Status Register structure.
Using the Subsystems
Status Subsystem 6-19
6
8
54 0
Operation Status Group
ST ATus : OPERa t i o n: CONDi t i o n?
STATus : OPERa t i on: EVENt ?
MSP
WFT MST
WFA
15
1632
256 64
Measurement stopped
Waiting for arming
Waiting for triggering
Measurement started
Figure 6-7 Bits in the Opeation Sta-
tus Register.

Table of Contents

Other manuals for Pendulum CNT-91R

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Pendulum CNT-91R and is the answer not in the manual?

Pendulum CNT-91R Specifications

General IconGeneral
BrandPendulum
ModelCNT-91R
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals