RF generator Pulse generator Counter
Frequency Power
Pulse
width
Pulse
repetitio
n
interval
Start
delay
Measu-
rement
time
Sensi-
tivity
Manual
acq
frequency
Power
measured
dBm
Tolerance P/F
1 GHz -15 dBm
100 µs
1 ms 10 µs 80 µs HIGH 1 GHz ±2 dBm
1 GHz -10 dBm
100 µs
1 ms 10 µs 80 µs HIGH 1 GHz
±2 dBm
1 GHz -5 dBm
100 µs
1 ms 10 µs 80 µs MED 1 GHz
±2 dBm
1 GHz -0 dBm
100 µs
1 ms 10 µs 80 µs LOW 1 GHz
±2 dBm
1 GHz +10 dBm 100 µs 1 ms 10 µs 80 µs LOW 1 GHz
±2 dBm
Table 7-19. Power sensitivity measurement
Performance Check
7-20
USER MANUAL ● CNT 9x Series ● Rev.22 February 2020
Power sensitivity:
P
ower accuracy:
Table 7-20. Power accuracy measurement
RF generator Pulse generator Counter
Frequency Power
Pulse
width
Pulse
repetition
interval
Start
delay
Measu-
rement
time
Sensi-
tivity
Manual
acq
frequency
Power
measured
Tolerance P/F
1 GHz -10 dBm 50 µs 1 ms 5µs 44 µs HIGH 1 GHz
±1.2 dBm
1 GHz -10 dBm 100 µs 1 ms 10 µs 80 µs HIGH 1 GHz
±1.2 dBm
1 GHz -10 dBm 1 ms 10 ms 100 µs 800 µs HIGH 1 GHz
±1.2 dBm

Table of Contents

Other manuals for Pendulum CNT-91R

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Pendulum CNT-91R and is the answer not in the manual?

Pendulum CNT-91R Specifications

General IconGeneral
BrandPendulum
ModelCNT-91R
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals