Specifications
Frequency & Period
Random Uncertainty (rms)
Smart Mode
For measuring times <200 ms and if Smart
Freq = AUTO or OFF:
x Measurement Result [Hz or s]
For measuring times >=200 ms and if Smart
Freq = AUTO or ON:
x Measurement Result [Hz or s]
N = 800 /Measuring Time, however, 6 < N <
1000 and
N < (Freq/2) * Measuring Time - 2 & N <
Measuring Time / 8 s
Normal Mode
x Measurement Result [Hz or s]
Systematic Uncertainty
Both Modes
Hz or
s]
where:
MR = Meas. Result (Freq or Per)
MT = Meas. Time
Frequency Ratio f
1
/f
2
Typical Random Uncertainty (rms)
NOTE. Frequency Ratio is an auxiliary mea-
surementfunction, intended to give an ind
ica-
tion, with no guaranteed specification.
[dimensionless, e.g. ppm]
Phase
Typical Random Uncertainty (rms)
NOTE. Phase is an auxiliary measurement
function, intended to give an indication, with no
guaranteed specification.
Systematic Uncertainty
Duty Factor
Random Uncertainty (rms)
x Freq.
or minimum: 1 ppm
Systematic Uncertainty
[dimensionless, e.g. ppm]
8-41
USER MANUAL ● CNT 9x Series ● Rev.22 February 2020

Table of Contents

Other manuals for Pendulum CNT-91R

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Pendulum CNT-91R and is the answer not in the manual?

Pendulum CNT-91R Specifications

General IconGeneral
BrandPendulum
ModelCNT-91R
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals