BG 5195 BEN / A (2011-05) TPG261.oi 73
Transmit: SEN [,x,x] <CR>[<LF>]
└─ Gauge 2, x =
0 –> No status change
1 –> Turn gauge off
2 –> Turn gauge on
└─ Gauge 1
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: x,x <CR><LF>
└─ Status gauge 2, x =
0 –> Gauge cannot be turned on/off
1 –> Gauge turned off
2 –> Gauge turned on
└─ Status gauge 1
Transmit: TID <CR>[<LF>]
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: x,x <CR><LF>
Identification gauge 2, x =
TPR (Pirani Gauge or
Pirani Capacitive gauge
1)
)
IKR9 (Cold Cathode Gauge 10
-9
)
IKR11 (Cold Cathode Gauge 10
-11
)
PKR (FullRange CC Gauge)
PBR (FullRange BA Gauge)
IMR (Pirani / High Pressure Gauge)
CMR (Linear gauge)
noSEn (no SEnsor)
noid (no identifier)
└─ Identification gauge 1
1)
TPR and PCR have identical identifiers.
There is no distinction made in communication
and in data evaluation, since pressure ranges
of these gauges are approximately the same.
Turning a gauge
on/off
Gauge identification